We use cookies to provide you with a top-notch browsing experience. You can find more information about cookies here. By selecting "Accept Cookies" you accept their placement.

Learn more

STR561, Digital panel meters
Indicator with strain-gauge input (for load cells and pressure sensors), 48x96mm. PIXSYS

Visualization, alarm and signal re-transmission functions, multi-voltage power supply, communication serial port
Product Inquiry

Product Inquiry

Indicator with strain-gauge input (for load cells and pressure sensors), 48x96mm. PIXSYS

    DETAILED DESCRIPTION

    STR561
    The indicator with strain-gauge input STR561 is designed as display for load cells and pressure sensors and includes isolated power supply for the strain-gauge. The conversion frequency programmable up to 1.2 KHz and the resolution of up to 24bit ensure high levels of accuracy. Various calibration modes can be selected for the analogue input (sample weight, percentage value on scale base, mV/V value). 
    Software key functions include multiple options for calibration management, Totalizer and Sum functions, graphic visualization of process trend with programmable time basis or bar- graphs with alarm/event thresholds. 
    An auto-configuration mode has also been implemented for selected Dynisco melt pressure sensors, simplifying the initial set-up for the user. Connectivity is guaranteed by  RS485 with Modbus RTU/Slave protocol. 
    For maximum flexibility of use, it is also possible to choose between horizontal or vertical installation.  
    Distinctive feature of the entire STR series is the innovative multilingual interface, with text menus allowing intuitive and quick navigation of parameters and visualization pages. 
    It is possible to choose among five languages and the comprehensive menu considerably reduces the need to consult technical manual for initial set-up.

    Specifications

    • Dimensions: 96x48 (Front panel) x 48 mm / or 1/8Din
    • Power supply24...230V AC/DC ±10% 50/60 Hz (galvanically isolated) 2500V
    • Power Consumption6 VA
    • Display: Oled (monochromatic yellow)
    • Operating conditionsTemperature 0-45 °C, humidity 35..95 RH% (non condensing)
    • Sealing Front panel:IP54 (IP65 with gasket) - Box and Terminal blocks: IP20
    • Quick set-up optionsMemory: Card microUSB, APP. MyPixsys (Andorid devices)
    • Wiring: extractable terminal blocks and spring locking
    • Visualization: range-999999 ...+999999
    • Analogue input: 1 x strain gauge Strain gauge input ADC 23 bit (8000000 points). Max 2 350Ω parallel load-cells or max 4 720Ω load-cells. Linearity < 0.01 % F.S. -thermal drift < 0.001% F.S. / °C . Max input signal 39 mV Max load-cell sensibility 7 mV/V. Potentiometer: min. 200 Ω (10000 points). Custom linearization up to 16 steps.
    • Digital inputs: 2 x digitalPNP inputs programmable for Run / Hold / Tare-Zero / Alarms manual reset / Peaks reset/ Min. peak reset / Activate-Reset Sum / Parameters lock. Net/gross weight visualization.
    • Sampling time: Configurable speed conversion from 1 Hz (8000000 points) to 1200 Hz (30000 points)
    • Relay outputs: 2 x relays 2A - 250VAC (resistive charge)
    • Analog output: 1 x analogue configurable as 0-20mA or 4-20mA. Resolution (16bit 60000 points)
    • Auxiliary output: 12 VDC - 30mA for digital inputs
    • Serial communication: RS485 Modbus RTU - Slave (galvanically isolated from Power supply/Inputs/Outputs)
    • Alarms regulation: ON - OFF with hysteresis
    • Alarm mode: Absolute / Threshold, Band with instantaneous/delayed/retentive/by digital input activation, sensor failure / Activation by serial line
    • Sum Function: By digital input or by keyboard it is possible to sum different process measurements over time
    • Totalizer Function: Visualisation of instant process value and total value since last reset
    • Trend visualization: Trend visualisation with selectable time basis 1 to 3600s
    • Analogue retransmission: Process values / Setpoints
    • Digital transmission via RS485: Process values / Setpoint / Parameters
    • Calibration function: Sampling value, full scale % value, value mV/V
    • Text menus: English/Italian/Deutsch/French/Spanish
    • Measure unit visualization: Selection of different measure units

    Ordering codes

    • STR561-12ABC-T128R: 1 Strain gauge analogue input + 2 relays 2A + 1 Out 0/4..20mA + RS485 Modbus RTU / Slave + Oled Display (monochromatic yellow)

    DOCUMENTS AND SPECIFICATIONS